Digi-Frame DF-1044r Manual de usuario Pagina 103

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 829
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 102
Triggers
R&S
®
RTO
87User Manual 1316.0827.02 ─ 06
RUN CONT. / RUN N×SINGLE
Front panel keys to start and stop a continuous acquisition or a defined number of acqui-
sition cycles, respectively. The number of acquisitions is set with "Average count".
SCPI command:
RUN on page 428
SINGle on page 428
STOP on page 428
Trigger out signal setup
Defines the pulse that is provided to the EXT TRIGGER OUT connector on the rear panel
when a trigger occurs.
To generate the trigger out signal, select "Enable trigger out".
"Polarity"
Sets the polarity of the trigger out pulse, that is the direction of the first
pulse edge.
"Pulse length"
Sets the length of the trigger out pulse.
"Delay"
Displays the delay of the first pulse edge to the trigger point. The delay
is always 250 ns.
SCPI command:
TRIGger<m>:OUT:STATe on page 491
TRIGger<m>:OUT:POLarity on page 491
TRIGger<m>:OUT:PLENgth on page 491
TRIGger<m>:OUT:DELay on page 491
Reference for Triggers
Vista de pagina 102
1 2 ... 98 99 100 101 102 103 104 105 106 107 108 ... 828 829

Comentarios a estos manuales

Sin comentarios