Remote Control
R&S
®
RTO
608User Manual 1316.0827.02 ─ 06
Parameters for setting and query:
<MaskTestName>
MTESt:ONViolation:STOP <MaskTestName>,<StopAcq>
MTESt:ONViolation:STOP? <MaskTestName>
Stops data acquisition for the specified event.
Parameters:
<StopAcq> NOACtion | VIOLation
See chapter 16.2.2.5, "Event Parameter", on page 421
*RST: NOACtion
Parameters for setting and query:
<MaskTestName>
MTESt:
ONViolation:PRINt <MaskTestName>,<Print>
MTESt:ONViolation:PRINt? <MaskTestName>
Prints a screenshot including the measurement results to the printer defined using
SYSTem:COMMunicate:PRINter:SELect<1..2> for the specified event.
Parameters:
<Print> NOACtion | SUCCess | VIOLation
See chapter 16.2.2.5, "Event Parameter", on page 421
*RST: NOACtion
Parameters for setting and query:
<MaskTestName>
MTESt:ONViolation:SAVewaveform <MaskTestName>,<SaveWfm>
MTESt:ONViolation:SAVewaveform? <MaskTestName>
Saves the waveform data.
Parameters:
<SaveWfm> NOACtion | SUCCess | VIOLation
See chapter 16.2.2.5, "Event Parameter", on page 421
*RST: NOACtion
Parameters for setting and query:
<MaskTestName>
16.2.12.5 Results
MTESt:RESult:STATe....................................................................................................609
MTESt:RESult[:RESult]...................................................................................................609
MTESt:RESult:COUNt:WAVeforms..................................................................................609
MTESt:RESult:COUNt:REMaining...................................................................................609
Command Reference