Remote Control
R&S
®
RTO
606User Manual 1316.0827.02 ─ 06
Parameters:
<Source> REFerence1 | REFerence2 | REFerence3 | REFerence4
*RST: REFerence1
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
Usage: SCPI confirmed
MTESt:WFMLupdate <MaskTestName>
Creates the upper and lower mask limit from the envelope of the selected reference
waveform. If the reference waveform was not defined before, it is created automatically
from the mask test source waveform which is set wtih MTESt:SOURce.
Setting parameters:
<MaskTestName> String containing the name of the mask test
Usage: Setting only
SCPI confirmed
MTESt:WFMRescale:XWIDth <MaskTestName>,<HorizontalWidth>
MTESt:WFMRescale:XWIDth? <MaskTestName>
Sets the width of the mask in horizontal direction. The specified factor in divisions is added
to the positive x-values and subtracted from the negative x-values of the mask limits in
relation to the source waveform of the mask.
Parameters:
<HorizontalWidth> Range: 0 to 100E+24
Increment: 0.05
*RST: 0
Default unit: div
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
MTESt:WFMRescale:YWIDth <MaskTestName>,<VerticalWidth>
MTESt:WFMRescale:YWIDth? <MaskTestName>
Sets the width of the waveform mask in vertical direction. The specified factor in divisions
is added to the y-values of the upper mask limit and subtracted from the y-values of the
lower mask limit. Thus, the upper half of the mask is pulled upwards, the lower half is
pulled down.
Parameters:
<VerticalWidth> Vertical mask width in divisions
Range: 0 to 100E+24
Increment: 0.05
*RST: 0
Default unit: div
Command Reference