Vista de pagina 534
Remote Control
R&S
®
RTO
519User Manual 1316.0827.02 ─ 06
Parameters:
<Coupling> ON | OFF
ON
Moving one cursor line moves the other cursor line too. The cursor
lines always remain a fixed distance.
OFF
Each cursor line is positioned independently.
*RST: OFF
CURSor<m>:
X1ENvelope <EnvelopeCurve1>
If the waveform arithmetics are set to envelope curve (see CHANnel<m>[:
WAVeform<n>]:ARIThmetics on page 434) and CURSor<m>:TRACking[:STATe]
is set to "ON", this setting defines how the first horizontal cursor is positioned.
Suffix:
<m>
.
1..4
math waveform
Parameters:
<EnvelopeCurve1> MIN | MAX | BOTH
MIN
The horizontal cursor is set to the crossing point of the vertical
cursor with the minimum waveform envelope.
MAX
The horizontal cursor is set to the crossing point of the vertical
cursor with the maximum waveform envelope.
BOTH
The envelope is ignored and the cursor is set to the crossing point
of the vertical cursor with the usual waveform
*RST: MAX
CURSor<m>:X2ENvelope <EnvelopeCurve2>
If the waveform arithmetics are set to envelope curve (see CHANnel<m>[:
WAVeform<n>]:ARIThmetics on page 434) and CURSor<m>:TRACking[:STATe]
is set to "ON", this setting defines how the second horizontal cursor is positioned.
Suffix:
<m>
.
1..4
math waveform