Remote Control
R&S
®
RTO
546User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..2
Selects the source number.
Parameters:
<EdgeCountDirection>FRFI | FRLA
FRFI - FRom FIrst, counting starts with the first edge of the wave-
form.
FRLA - FRom LAst, counting starts with the last edge of the wave-
form.
*RST: FRFI
MEASurement<m>:
AMPTime:DELay<n>:ECOunt <EdgeIndex>
Sets the number of the edge that is relevant for delay measurement for each source.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..2
Selects the source number.
Parameters:
<EdgeIndex> Edge number
Range: 1 to 100000
Increment: 1
*RST: 1
MEASurement<m>:AMPTime:DELay<n>:LSELect <DelayLevSelection>
Selects the reference level on which the time is measured for each source.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..2
Selects the source number.
Parameters:
<DelayLevSelection> UPPer | MIDDle | LOWer
*RST: MIDDle
MEASurement<m>:AMPTime:DELay<n>:SLOPe <Slope>
Sets the edge of each source, between which the delay is measured.
Command Reference