Digi-Frame DF-1044r Manual de usuario Pagina 551

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 829
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 550
Remote Control
R&S
®
RTO
535User Manual 1316.0827.02 ─ 06
REFLevel<m>:ABSolute:MLEVel <MiddleLevel>
The middle reference level.
Suffix:
<m>
.
2...21
Source waveform, see chapter 16.2.2.1, "Waveform Suf-
fix", on page 419.
Parameters:
<MiddleLevel> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Example:
REFLevel2:ABSolute:MLEVel 0.005
Sets the middle reference level for Channel1/Waveform1 to 5 mV.
C1W1 corresponds to suffix number 2.
See also: example "Manual reference level definition using abso-
lute values" on page 527
Usage: SCPI confirmed
REFLevel<m>:ABSolute:LLEVel <LowerLevel>
The lower reference level, required e.g. to determine a fall.
Suffix:
<m>
.
2...21
Source waveform, see chapter 16.2.2.1, "Waveform Suf-
fix", on page 419.
Parameters:
<LowerLevel> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Example:
REFLevel2:ABSolute:LLEVel 0.001
Sets the lower reference level for Channel1/Waveform1 to 1 mV.
C1W1 corresponds to suffix number 2.
See also: example "Manual reference level definition using abso-
lute values" on page 527
Usage: SCPI confirmed
REFLevel<m>:RELative:UPPer <UppRefLevRel>
Sets the upper relative reference level if REFLevel<m>:RELative:MODE is set to
USER.
Command Reference
Vista de pagina 550
1 2 ... 546 547 548 549 550 551 552 553 554 555 556 ... 828 829

Comentarios a estos manuales

Sin comentarios