Remote Control
R&S
®
RTO
704User Manual 1316.0827.02 ─ 06
Parameters:
<Type> SYNC | ID | IDOR | IDDT | WKFR | ERRC
SYNC
Start of the frame, triggers on the stop bit of the sync field.
ID
Sets the trigger to one specific identifier or an identifier range.
To set the identifier, use TRIGger<m>:LIN:ICONdition,
TRIGger<m>:LIN:IMIN on page 705, and TRIGger<m>:
LIN:IMAX on page 705.
IDOR
Triggers on an OR combination with up to four four identifier con-
ditions. For each identifier condition, enable it with
TRIGger<m>:LIN:IDOR<n>:ENABle and set the value with
TRIGger<m>:LIN:IDOR<n>[:VALue]
IDDT
Combination of identifier and data conditions
To set the identifier condition, use TRIGger<m>:LIN:
ICONdition, TRIGger<m>:LIN:IMIN, and TRIGger<m>:
LIN:IMAX.
To set the data condition, use TRIGger<m>:LIN:DCONdition,
TRIGger<m>:LIN:DMIN, and TRIGger<m>:LIN:DMAX.
WKFR
Wakeup frame
ERRC
Error condition. Define the error types with TRIGger<m>:LIN:
CHKSerror on page 709, TRIGger<m>:LIN:IPERror, and
TRIGger<m>:LIN:SYERror
*RST:
SYNC
TRIGger<m>:LIN:ICONdition <IdentifierOperator>
Sets the operator to set a specific identifier or an identifier range.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only