Remote Control
R&S
®
RTO
587User Manual 1316.0827.02 ─ 06
Parameters:
<Range> Range: 1 to 500
Increment: 1
*RST: 100
Default unit: dB
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:MAGNitude:SCALe <MagnitudeScale>
Defines the scaling of the y-axis. The display values are valid for 50Ω termination impend-
ance.
For details on the available scaling modes, see "Magnitude unit" on page 203.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<MagnitudeScale> LINear | DBM | DB | DBUV | DBMV | DBV
LINear
linear scaling; displays the RMS value of the voltage
DB
logarithmic scaling; related to reference level
*RST: DBM
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:PHASe:SCALe <PhaseScale>
Defines the scaling unit for phase display.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<PhaseScale> DEGRees | RADians
*RST: DEGRees
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:PHASe:SUPPression <Suppression>
Enables noise suppression. Phase calculation is restricted to frequencies with a minimum
magnitude, the threshold value (see CALCulate:MATH<m>:FFT:PHASe:THReshold
on page 588).
Suffix:
<m>
.
1...4
math waveform
Command Reference