Remote Control
R&S
®
RTO
425User Manual 1316.0827.02 ─ 06
Parameters:
<Action> 0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL <Number>
ReCaLl calls up the instrument settings from an intermediate memory identified by the
specified number. The instrument settings can be stored to this memory using the com-
mand *SAV with the associated number.
It also activates the instrument settings which are stored in a file and loaded using
MMEMory:LOAD:STATe.
*RST
ReSeT
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
Usage:
Setting only
*SAV <Number>
SAVe stores the current instrument settings under the specified number in an intermedi-
ate memory. The settings can be recalled using the command *RCL with the associated
number.
To transfer the stored instrument settings to a file, use MMEMory:STORe:STATe.
*SRE <Contents>
Service Request Enable
Sets the service request enable register to the indicated value. This command determines
under which conditions a service request is triggered.
Parameters:
<Contents> Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range: 0 to 255
*STB?
STatus Byte query
Reads the contents of the status byte in decimal form.